About 1 in 6 med spa websites we checked misstate a cleared device's FDA status

5 min readStetwright Research

14 of 87 med spa and aesthetic clinic sites misstated the FDA status of a device that holds only a clearance or a De Novo grant.

We do not name any business. Figures are counts across the sites we read, and nothing here is a legal finding about any site.

Contents

Summary

14 of 87 med spa and aesthetic clinic websites we checked (16%, 95% interval 10% to 25%) described a device as "FDA-approved" when the FDA record shows only a 510(k) clearance or a De Novo grant. Counting device pages our earlier reads found and our crawl did not reach, the figure rises to 23 of 87. The same sites rarely told self-pay patients they can ask for a good faith estimate: 81 of 83 showed no notice of that right on the pages we read.

Why the word matters

The FDA uses different words for different reviews. A device that is shown to be substantially equivalent to one already on the market gets a 510(k) clearance. A new low-to-moderate risk device can get a De Novo grant. "Approval" is the word the FDA uses for premarket approval (PMA) and humanitarian device exemptions, and its device listings describe PMA as "a more rigorous premarket review than the 510(k) pathway."

For 510(k) devices, the FDA's rule at 21 CFR 807.97 says that a premarket notification and the FDA's finding of substantial equivalence "does not in any way denote official approval of the device," and that "any representation that creates an impression of official approval of a device because of complying with the premarket notification regulations is misleading and constitutes misbranding."

A patient who reads "FDA-approved" may assume the device went through the FDA's most demanding review, which none of the devices in the counted claims had.

What we found

Measure Sites Rate 95% interval
Called a cleared or De Novo device "FDA-approved" (crawl, hand-checked) 14 of 87 16% 10% to 25%
Same, seen on two separate reads 12 of 87 14% 8% to 23%
Adding approval wording our Nov 20 reads found on pages the crawl missed 23 of 87 26% 18% to 37%

The crawl figure is a floor. The crawl read a fixed set of pages on each site and often missed the device pages. Our Nov 20 page reads found approval wording on 9 more of the 87 sites. Two of those 9 used it in history statements about how a device came to market, and one gave approval wording to a laser we could not find in the FDA's databases. Dropping the two history statements gives 21 of 87 (24%, 16% to 34%).

The claims named devices with an FDA record. In the crawl, no med spa page claimed FDA status for a device with no FDA record (0 of 87). The one such case came from our Nov 20 reads. On those reads, no checkable claim tied FDA status to a use clearly outside the device's cleared indications (0 of 55 claims).

Med spas stood out from the rest of the panel. We ran the same FDA-status check on 785 readable healthcare sites in ten segments. Hand review confirmed a misstatement on 22 (2.8%, 1.9% to 4.2%), and med spas accounted for 14 of the 22. Hospitals (0 of 433) and rehab programs (0 of 55) had none, though the automated flag fired on 9 hospital sites and 7 rehab sites. Most flags across the panel were correct statements, such as approvals of brand-name drugs or notes that compounded drugs are not FDA-approved.

Good faith estimates

The No Surprises Act rule at 45 CFR 149.610(b)(1)(iii)(A) says information on the availability of a good faith estimate for uninsured or self-pay patients is to be "prominently displayed (and easily searchable from a public search engine) on the convening provider's or convening facility's website."

Measure Sites Rate 95% interval
No notice of the good faith estimate right on the pages read 81 of 83 98% 92% to 99%
Same, counting a footer link to a No Surprises Act page as a notice 77 of 83 93% 85% to 97%

Whether a given cosmetic service falls within the rule is a legal question we do not answer here.

What a patient can do

Ask the clinic for the device's brand and model, then look it up in the FDA's 510(k), De Novo and PMA databases at accessdata.fda.gov. The record shows which review the device went through and the uses it was cleared or approved for. If you are paying yourself, ask for a good faith estimate before you book.

Method

  • Sites. The med spa and aesthetic segment of a fixed 895-site healthcare panel, built from fixed searches for "med spa" and "body contouring" in the 25 largest US metro areas. 88 sites had readable pages; 87 could be judged for the FDA check and 83 for the good faith estimate check. A separate Nov 20 read covered 60 sites drawn in a fixed random order from the same searches, most of which are also in the panel.
  • Dates. Page reads on Nov 20, 2026. Panel crawl Nov 20 to 21, 2026. Hand review and re-reads Nov 21.
  • FDA records. Each device named in an approval claim was looked up in the FDA's 510(k), De Novo and PMA records. A claim was counted only when the record showed a 510(k) clearance or De Novo grant and no PMA.
  • Two reads. A positive counts as confirmed once seen two ways: in the crawl plus either the Nov 20 read or a Nov 21 re-read. 12 of the 14 crawl positives meet that test. The 9 additions in the 23 of 87 figure rest on the Nov 20 reads alone.
  • What "not found" means. Not seen on the pages read. The crawl read about 3 to 16 pages per site, stored short excerpts around each match, and could not see text in collapsed sections or page titles. It ran no site search.
  • Statistics. Rates use Wilson 95% intervals. The unit is the site.
  • Scope. We report what each page says and what the FDA record and rule text say. Nothing here is a legal finding about any clinic, and no clinic is named.

Sources: FDA, 21 CFR 807.97 (eCFR); FDA, Recently-Approved Devices (PMA, 510(k), HDE and De Novo listings); FDA, "Is It Really 'FDA Approved'?"; 45 CFR 149.610 (eCFR).

A business that wants to check its own site for issues like these can request a Stetwright website review.

More healthcare website studies

All studies

Next step

Check your own site against the same rules.

A full-site review reads your pages against FDA, FTC and state rules and shows each finding with its rule and source.